History of the oscilloscope

Illustration of Joubert's step-by-step method of hand-plotting waveform measurements[1]


The history of the oscilloscope was fundamental to science because an oscilloscope is a device for viewing waveform oscillations, as of electrical voltage or current, in order to measure frequency and other wave characteristics. This was important in developing electromagnetic theory. The first recordings of waveforms were with a galvanometer coupled to a mechanical drawing system dating from the second decade of the 19th century. The modern day digital oscilloscope is a consequence of multiple generations of development of the oscillograph, cathode-ray tubes, analog oscilloscopes, and digital electronics.

Hand-drawn oscillograms

The earliest method of creating an image of a waveform was through a laborious and painstaking process of measuring the voltage or current of a spinning rotor at specific points around the axis of the rotor, and noting the measurements taken with a galvanometer. By slowly advancing around the rotor, a general standing wave can be drawn on graphing paper by recording the degrees of rotation and the meter strength at each position.

This process was first partially automated by Jules François Joubert [fr] with his step-by-step method of wave form measurement. This consisted of a special single-contact commutator attached to the shaft of a spinning rotor. The contact point could be moved around the rotor following a precise degree indicator scale and the output appearing on a galvanometer, to be hand-graphed by the technician.[2] This process could only produce a very rough waveform approximation since it was formed over a period of several thousand wave cycles, but it was the first step in the science of waveform imaging.

Automatic paper-drawn oscillograph

Schematic and perspective view of the Hospitalier Ondograph, which used a pen on a paper drum to record a waveform image built up over time, using a synchronous motor drive mechanism and a permanent magnet galvanometer[3][4]

The first automated oscillographs used a galvanometer to move a pen across a scroll or drum of paper, capturing wave patterns onto a continuously moving scroll. Due to the relatively high-frequency speed of the waveforms compared to the slow reaction time of the mechanical components, the waveform image was not drawn directly but instead built up over a period of time by combining small pieces of many different waveforms, to create an averaged shape.

The device known as the Hospitalier Ondograph was based on this method of wave form measurement. It automatically charged a capacitor from each 100th wave, and discharged the stored energy through a recording galvanometer, with each successive charge of the capacitor being taken from a point a little farther along the wave.[5] (Such wave-form measurements were still averaged over many hundreds of wave cycles but were more accurate than hand-drawn oscillograms.)

Tilting mirror

Duddell moving-coil oscillograph with mirror and two supporting moving coils on each side of it[6]
Rotating shutter and moving mirror assembly for placing time-index marks next to the waveform pattern[7]
Moving-film camera for recording the waveform[8]

In order to permit direct measurement of waveforms it was necessary for the recording device to use a very low-mass measurement system that can move with sufficient speed to match the motion of the actual waves being measured. This reduced the measurement device to a small mirror that could move at high speeds to match the waveform.

The term "oscillograph" was coined by André Blondel in 1893 to refer to his instrument based on the earlier-known mirror galvanometer but adapted to recording high-frequency oscillations.[9] William Duddell also developed a similar instrument few years later.

Film recording of sparking across switch contacts, as a high-voltage circuit is disconnected[10]

To perform a waveform measurement, a photographic slide would be dropped past a window where the light beam emerges, or a continuous roll of motion picture film would be scrolled across the aperture to record the waveform over time. Although the measurements were much more precise than the built-up paper recorders, there was still room for improvement due to having to develop the exposed images before they could be examined.

The term "oscilloscope" was coined in 1907.[11]

In the 1920s, a tiny tilting mirror attached to a diaphragm at the apex of a horn provided good response up to a few kHz, perhaps even 10 kHz. A time base, unsynchronized, was provided by a spinning mirror polygon, and a collimated beam of light from an arc lamp projected the waveform onto the lab wall or a screen.[12]

Even earlier, audio applied to a diaphragm on the gas feed to a flame made the flame height vary, and a spinning mirror polygon gave an early glimpse of waveforms.[13]

Moving-paper oscillographs using UV-sensitive paper and advanced mirror galvanometers provided multi-channel recordings in the mid-20th century. Frequency response was into at least the low audio range.

CRT invention

The interior of a cathode-ray tube for use in an oscilloscope. 1. Deflection voltage electrode; 2. Electron gun; 3. Electron beam; 4. Focusing coil; 5. Phosphor-coated inner side of the screen

Cathode ray tubes (CRTs) were developed in the late 19th century. At that time, the tubes were intended primarily to demonstrate and explore the physics of electrons (then known as cathode rays). Karl Ferdinand Braun invented the CRT oscilloscope as a physics curiosity in 1897, by applying an oscillating signal to electrically charged deflector plates in a phosphor-coated CRT. Braun tubes were laboratory apparatus, using a cold-cathode emitter and very high voltages (on the order of 20,000 to 30,000 volts). With only vertical deflection applied to the internal plates, the face of the tube was observed through a rotating mirror to provide a horizontal time base.[14] In 1899 Jonathan Zenneck equipped the cathode ray tube with beam-forming plates and used a magnetic field for sweeping the trace.[15]

Early cathode ray tubes had been applied experimentally to laboratory measurements as early as 1919 [16] but suffered from poor stability of the vacuum and the cathode emitters. The application of a thermionic emitter allowed operating voltage to be dropped to a few hundred volts. Western Electric introduced a commercial tube of this type, which relied on a small amount of gas within the tube to assist in focusing the electron beam.[16]

V. K. Zworykin described a permanently sealed, high-vacuum cathode ray tube with a thermionic emitter in 1931. This stable and reproducible component allowed General Radio to manufacture an oscilloscope that was usable outside a laboratory setting.[15]

The first dual-beam oscilloscope was developed in the late 1930s by the British company A.C.Cossor (later acquired by Raytheon). The CRT was not a true double beam type but used a split beam made by placing a third plate between the vertical deflection plates. It was widely used during WWII for the development and servicing of radar equipment. Although extremely useful for examining the performance of pulse circuits it was not calibrated so could not be used as a measuring device. It was, however, useful in producing response curves of IF circuits and consequently a great aid in their accurate alignment.

Allen B. Du Mont Labs. made moving-film cameras, in which continuous film motion provided the time base. Horizontal deflection was probably disabled, although a very slow sweep would have spread phosphor wear. CRTs with P11 phosphor were either standard or available.[17]

Long-persistence CRTs, sometimes used in oscilloscopes for displaying slowly changing signals or single-shot events, used a phosphor such as P7, which comprised a double layer. The inner layer fluoresced bright blue from the electron beam, and its light excited a phosphorescent "outer" layer, directly visible inside the envelope (bulb). The latter stored the light, and released it with a yellowish glow with decaying brightness over tens of seconds. This type of phosphor was also used in radar analog PPI CRT displays, which are a graphic decoration (rotating radial light bar) in some TV weather-report scenes.

Sweep circuit

Oscilloscope with synchronized sweep. "HOR. SELECTOR" sets horizontal frequency range (the capacitor); "FREQ. VERNIER" adjusts the free-running frequency; "SYNC. AMPLITUDE" sets the gain to the comparator

The technology for the horizontal sweep, that portion of the oscilloscope that creates the horizontal time axis, has changed.

Synchronized sweep

Early oscilloscopes used a synchronized sawtooth waveform generator to provide the time axis. The sawtooth would be made by charging a capacitor with a relatively constant current; that would create a rising voltage. The rising voltage would be fed to the horizontal deflection plates to create the sweep. The rising voltage would also be fed to a comparator; when the capacitor reached a certain level, the capacitor would be discharged, the trace would return to the left, and the capacitor (and the sweep) would start another traverse. The operator would adjust the charging current so the sawtooth generator would have a slightly longer period than a multiple of the vertical axis signal. For example, when looking at a 1 kHz sinewave (1 ms period), the operator might adjust the horizontal frequency to a little bit more than 5 ms. When the input signal was absent, the sweep would free run at that frequency.

If the input signal were present, the resulting display would not be stable at the horizontal sweep's free-running frequency because it was not a submultiple of the input (vertical axis) signal. To fix that, the sweep generator would be synchronized by adding a scaled version of the input signal to the sweep generator's comparator. The added signal would cause the comparator to trip a little earlier and thus synchronize it to the input signal. The operator could adjust the sync level; for some designs, the operator could choose the polarity.[18] The sweep generator would turn off (known as blanking) the beam during retrace.[19]

The resulting horizontal sweep speed was uncalibrated because the sweep rate was adjusted by changing slope of the sawtooth generator. The time per division on the display depended upon the sweep's free-running frequency and a horizontal gain control.

A synchronized sweep oscilloscope could not display a non-periodic signal because it could not synchronize the sweep generator to that signal. Horizontal circuits were often AC-coupled

Triggered sweep

Triggered sweep controls on a Tektronix 465 oscilloscope

During World War II, a few oscilloscopes used for radar development (and a few laboratory oscilloscopes) had so-called driven sweeps. These sweep circuits remained dormant, with the CRT beam cut off, until a drive pulse from an external device unblanked the CRT and started a constant-speed horizontal trace; the calibrated speed permitted measurement of time intervals. When the sweep was complete, the sweep circuit blanked the CRT (turned off the beam), reset itself, and waited for the next drive pulse. The Dumont 248, a commercially available oscilloscope produced in 1945, had this feature.

Oscilloscopes became a much more useful tool in 1946 when Howard Vollum and Melvin Jack Murdock introduced the Tektronix Model 511 triggered-sweep oscilloscope. Howard Vollum had first seen this technology in use in Germany. The triggered sweep has a circuit that develops the driven sweep's drive pulse from the input signal.

Triggering allows stationary display of a repeating waveform, as multiple repetitions of the waveform are drawn over exactly the same trace on the phosphor screen. A triggered sweep maintains the calibration of sweep speed, making it possible to measure properties of the waveform such as frequency, phase, rise time, and others, that would not otherwise be possible.[20] Furthermore, triggering can occur at varying intervals, so there is no requirement that the input signal be periodic.

Triggered-sweep oscilloscopes compare the vertical deflection signal (or rate of change of the signal) with an adjustable threshold, referred to as trigger level. As well, the trigger circuits also recognize the slope direction of the vertical signal when it crosses the threshold—whether the vertical signal is positive-going or negative-going at the crossing. This is called trigger polarity. When the vertical signal crosses the set trigger level and in the desired direction, the trigger circuit unblanks the CRT and starts an accurate linear sweep. After the completion of the horizontal sweep, the next sweep will occur when the signal once again crosses the threshold trigger.

Variations in triggered-sweep oscilloscopes include models offered with CRTs using long-persistence phosphors, such as type P7. These oscilloscopes were used for applications where the horizontal trace speed was very slow, or there was a long delay between sweeps, to provide a persistent screen image. Oscilloscopes without triggered sweep could also be retro-fitted with triggered sweep using a solid-state circuit developed by Harry Garland and Roger Melen in 1971.[21]

As oscilloscopes have become more powerful over time, enhanced triggering options allow capture and display of more complex waveforms. For example, trigger holdoff is a feature in most modern oscilloscopes that can be used to define a certain period following a trigger during which the oscilloscope will not trigger again. This makes it easier to establish a stable view of a waveform with multiple edges which would otherwise cause another trigger.[citation needed]

Tektronix

Type 465 Tektronix oscilloscope, a popular analog oscilloscope during the 1980s

Vollum and Murdock went on to found Tektronix, the first manufacturer of calibrated oscilloscopes (which included a graticule on the screen and produced plots with calibrated scales on the axes of the screen).[citation needed] Later developments by Tektronix included the development of multiple-trace oscilloscopes for comparing signals either by time-multiplexing (via chopping or trace alternation) or by the presence of multiple electron guns in the tube. In 1963, Tektronix introduced the Direct View Bistable Storage Tube (DVBST), which allowed observing single pulse waveforms rather than (as previously) only repeating wave forms. Using micro-channel plates, a variety of secondary-emission electron multiplier inside the CRT and behind the faceplate, the most advanced analog oscilloscopes (for example, the Tek 7104 mainframe) could display a visible trace (or allow the photography) of a single-shot event even when running at extremely fast sweep speeds. This oscilloscope went to 1 GHz.

In vacuum-tube oscilloscopes made by Tektronix, the vertical amplifier's delay line was a long frame, L-shaped for space reasons, that carried several dozen discrete inductors and a corresponding number of low-capacitance adjustable ("trimmer") cylindrical capacitors. These oscilloscopes had plug-in vertical input channels. For adjusting the delay line capacitors, a high-pressure gas-filled mercury-wetted reed switch created extremely fast-rise pulses which went directly to the later stages of the vertical amplifier. With a fast sweep, any misadjustment created a dip or bump, and touching a capacitor made its local part of the waveform change. Adjusting the capacitor made its bump disappear. Eventually, a flat top resulted.

Vacuum-tube output stages in early wideband oscilloscopes used radio transmitting tubes, but they consumed a lot of power. Picofarads of capacitance to ground limited bandwidth. A better design, called a distributed amplifier, used multiple tubes, but their inputs (control grids) were connected along a tapped L-C delay line, so the tubes' input capacitances became part of the delay line. As well, their outputs (plates/anodes) were likewise connected to another tapped delay line, its output feeding the deflection plates. This amplifier was often push-pull, so there were four delay lines, two for input (grid), and two for output (plate).

Digital oscilloscopes

The first digital storage oscilloscope (DSO) was built by Nicolet Test Instrument of Madison, Wisconsin.[citation needed] It used a low-speed analog-to-digital converter (1 MHz, 12 bit) used primarily for vibration and medical analysis.[citation needed] The first high-speed DSO (100 MHz, 8 bit) was developed by Walter LeCroy, who founded the LeCroy Corporation of New York, USA, after producing high-speed digitizers for the research center CERN in Switzerland. LeCroy (since 2012 Teledyne LeCroy) remains one of the three largest manufacturers of oscilloscopes in the world.[citation needed]

Starting in the 1980s, digital oscilloscopes became prevalent. Digital storage oscilloscopes use a fast analog-to-digital converter and memory chips to record and show a digital representation of a waveform, yielding much more flexibility for triggering, analysis, and display than is possible with a classic analog oscilloscope. Unlike its analog predecessor, the digital storage oscilloscope can show pre-trigger events, opening another dimension to the recording of rare or intermittent events and troubleshooting of electronic glitches. As of 2006 most new oscilloscopes (aside from education and a few niche markets) are digital.

Digital scopes rely on effective use of the installed memory and trigger functions: not enough memory and the user will miss the events they want to examine; if the scope has a large memory but does not trigger as desired, the user will have difficulty finding the event.

DSOs also led to the creation of hand-held digital oscilloscopes, useful for many test and field service applications. A hand held oscilloscope is usually a real-time oscilloscope, using a monochrome or color liquid crystal display for its display.

Due to the rise in the prevalence of PCs, PC-based oscilloscopes have been becoming more common. The PC platform can be part of a standalone oscilloscope or as a standalone PC in combination with an external oscilloscope. With external oscilloscopes, a signal will be captured on external hardware (which includes an analog-to-digital converter and memory) and transmitted to the computer, where it is processed and displayed.

Notes

  1. ^ Hawkins (1917, p. 1844) Fig. 2589
  2. ^ Hawkins (1917, pp. 1841–1846)
  3. ^ Hawkins (1917, p. 1850), Fig. 2597
  4. ^ Hawkins (1917, p. 1851), Fig. 2598
  5. ^ Hawkins (1917, pp. 1849–1851)
  6. ^ Hawkins (1917, p. 1858), Fig. 2607
  7. ^ Hawkins (1917, p. 1855), Fig. 2620
  8. ^ Hawkins (1917, p. 1866), Figs. 2621–2623
  9. ^ Chisholm, Hugh, ed. (1911). "1911 Encyclopædia Britannica/Oscillograph" . Encyclopædia Britannica (11th ed.). Cambridge University Press.
  10. ^ Hawkins (1917, p. 1867), Fig. 2625
  11. ^ "oscilloscope | Etymology of oscilloscope by etymonline". www.etymonline.com. Retrieved 2024-10-06.
  12. ^ The Science of Musical Sounds, by Dayton Clarence Miller, published about 1924
  13. ^ Late 1800s, possibly? Ganot's Physics, a very popular 19th-C physics text? IIrc, Miller also described this.
  14. ^ Abramson (1995, p. 13)
  15. ^ a b Kularatna, Nihal (2003). "Chapter 5: Fundamentals of Oscilloscopes". Digital and analogue instrumentation: testing and measurement. Institution of Engineering and Technology. p. 165. ISBN 978-0-85296-999-1. Retrieved 2011-01-19.
  16. ^ a b Burns (1998, pp. 346–347)
  17. ^ Oscilloscope catalog, Allen B. DuMont Labs, possibly 1949
  18. ^ Operator's Manual: Model KG-635 DC to 5.2 MC 5" Wideband Oscilloscope, Maywood, IL: Knight Electronics Corporation, 1965, p. 3, Synchronization ... + internal, − internal, 60 cps, and external. Sync limiting provides semi-automatic operation with level control. Locks from waveform fundamentals up to 5 mc. Will sync on display amplitudes as low as 0.1 [inch] The KG-635 sync amplifier used a 12AT7 differential amplifier (V5). (id p. 15.) Sync level control would bias the amplifier into cutoff so action would only occur near the end of the sweep; the sync output was a negative pulse to the sweep generator; a diode pulse limiter clamped the sync pulse. (id p. 18.)
  19. ^ KG-635 p. 18 stating, "Retrace blanking is obtained from the plate of V-6A and applied to the cathode of the CRT."
  20. ^ Spitzer & Howarth 1972, p. 122
  21. ^ Garland, Harry; Melen, Roger (1971). "Add Triggered Sweep to your Scope". Popular Electronics. 35 (1): 61–66.

References

Read other articles:

The Tortoise and the Hare Serial Silly SymphoniesBerkas:Tobymax.jpgToby Tortoise dan Max HareSutradaraWilfred JacksonProduserWalt DisneyIde CeritaLarry ClemmonsMusikFrank ChurchillAnimatorHamilton LuskeLarry ClemmonsDick LundyLes ClarkLouie SchmittDick HuemerEric LarsonFrenchy de TremaudanMilt SchafferStudioWalt Disney ProductionsDidistribusikan olehUnited ArtistsDikeluarkan pada date(s) 5 Januari 1935 (1935-01-05) Format warnaTechnicolorLama waktu8 menit 17 detikBahasaInggris The Tortoi...

 

Ada LovelaceLahirThe Hon. Augusta Ada Byron (1815-12-10)10 Desember 1815London, InggrisMeninggal27 November 1852(1852-11-27) (umur 36)Marylebone, London, InggrisMakamGereja St. Mary Magdalene, Hucknall, NottinghamKebangsaanBritishGelarCountess LovelaceSuami/istriWilliam King-Noel, Earl pertama dari LovelaceAnak Byron King-Noel, Viscount Ockham dan Baron Wentworth ke-12 Anne Blunt, Baroness Wentworth ke-15 Ralph King-Milbanke, Earl ke-2 dari Lovelace Augusta Ada King, Countess Lovelace (1...

 

الدوري الهولندي الممتاز تفاصيل الموسم 1956-1957 النسخة 1  البلد هولندا  التاريخ بداية:2 سبتمبر 1956  نهاية:2 يونيو 1957  المنظم الاتحاد الملكي الهولندي لكرة القدم  البطل أياكس أمستردام مباريات ملعوبة 306   عدد المشاركين 18   أهداف مسجلة 1139   - 1957-1958 تعديل مصدري - تعديل...

Dieser Artikel behandelt den Fernsehsender, für den Radiosender SRF 2 siehe Radio SRF 2 Kultur. SRF zwei Fernsehsender (privatrechtlich im Auftrag des Bundes) Programmtyp Vollprogramm Empfang digital terrestrisch, Kabel, Satellit Bildauflösung Full HD (1080p) bei 50 Hz Sendestart 1. Sep. 1997 Sitz Zürich, Schweiz Sendeanstalt Schweizer Radio und Fernsehen Intendant Nathalie Wappler Programmchef Hansruedi Schoch SRF zwei ist das zweite Fernsehprogramm von Schweizer Radio und Fernsehen ...

 

Untuk kegunaan lain, lihat Memoir of a Murderer (disambiguasi). Memoir of a MurdererPoster rilis teatrikalNama lain살인자의 기억법Sutradara Won Shin-yun Produser You Jeong-hun Won Shin-yun Ditulis oleh Hwang Jo-yun Won Shin-yun SkenarioHwang Jo-yunWon Shin-yunBerdasarkanA Murderer's Guide to Memorizationoleh Kim Young-haPemeranSol Kyung-guKim Nam-gilKim Seol-hyunOh Dal-suDistributorShowbox/MediaplexTanggal rilis 06 September 2017 (2017-09-06) Durasi118 menit128 menit (pemoto...

 

Академія медичних наук СРСР Основні дані Засновано 1944 Країна  СРСРТип академія наук і організаціяМатеринськаорганізація Академія наук СРСР Акаде́мія меди́чних нау́к СРСР — вища наукова медична установа в СРСР. Створена в 1944 році. ЇЇ приємниками є Російська А...

1467 battle between the republics of Venice and Florence This article does not cite any sources. Please help improve this article by adding citations to reliable sources. Unsourced material may be challenged and removed.Find sources: Battle of Molinella – news · newspapers · books · scholar · JSTOR (November 2012) (Learn how and when to remove this template message) Battle of MolinellaDepiction of the battle in the Malpaga CastleDateJuly 25, 1467Locati...

 

This article needs additional citations for verification. Please help improve this article by adding citations to reliable sources. Unsourced material may be challenged and removed.Find sources: Rosario Tijeras film – news · newspapers · books · scholar · JSTOR (April 2017) (Learn how and when to remove this template message) 2005 filmRosario TijerasTheatrical release posterDirected byEmilio MailléScreenplay byMarcelo FiguerasBased onRosario Tije...

 

  لمعانٍ أخرى، طالع جو أندرسون (توضيح). جو اندرسون معلومات شخصية الميلاد 26 مارس 1982 (العمر 41 سنة)إنجلترا مواطنة المملكة المتحدة  الأب مايلز اندرسون  الحياة العملية المدرسة الأم أكاديمية ويبر دوغلاس للفنون الدرامية  [لغات أخرى]‏  المهنة ممثل اللغات الإنجل...

American politician (1849–1936) For other people with the same name, see Hamilton Fish (disambiguation). Hamilton Fish IIMember of the U.S. House of Representativesfrom New York's 21st districtIn officeMarch 4, 1909 – March 3, 1911Preceded bySamuel McMillanSucceeded byRichard E. ConnellAssistant Treasurer of the United StatesIn office1903–1908Preceded byConrad N. JordanSucceeded byGeorge S. TerrySpeaker of the New York State AssemblyIn office1895–1896Preceded byGe...

 

American space robotics company Astrobotic Technology, Inc.TypePrivate companyIndustryAerospaceRoboticsFounded2007; 16 years ago (2007)HeadquartersPittsburgh, Pennsylvania, U.S.Key peopleJohn Thornton (CEO)Red Whittaker (Chairman)Paul C. O'Brien (Board of Directors)William Readdy (Board of Directors)Products Autonomous robots space systems space payloads robotic spacecraft telerobotics Number of employees50–100 (As of July 1, 2012[update])Websitewww.astrob...

 

Canadian restaurant chain This article is about the restaurant chain. For the physical room of a house, see Recreation room. For the video game, see Rec Room (video game). For the fictional holodeck-like device, see The Practical Joker. The Rec RoomTypeSubsidiaryIndustryEntertainment restaurantsFounded2016 in EdmontonNumber of locations10ParentCineplex EntertainmentWebsitetherecroom.com The REC Room in Square One, Mississauga The Rec Room is a Canadian chain of entertainment restaurants owned...

Jutland DivisionActive1 Jun 1952 - 31 Dec 1996CountryDenmarkRoleMechanized infantrySizeDivisionPart ofAllied Forces Baltic Approaches AFNORTHWestern Regional CommandMilitary unit The Jutland Division (Jyske Divisionskommando) was a mechanized infantry formation of the Danish Army during the Cold War. It was formed during 1951-1952 as 3rd Division from assets of the III Military Region and the division headquarters was initially located in Aabenraa. As initially formed, the division had o...

 

Town in Irbid Governorate, JordanAl Mazar al Shamali المزار الشماليTownAl Mazar al ShamaliLocation in JordanCoordinates: 32°28′21″N 35°47′34″E / 32.47250°N 35.79278°E / 32.47250; 35.79278Country JordanGovernorateIrbid GovernoratePopulation • Estimate ()16,000Time zoneUTC + 2 Al-Mazar Al-Shamali or Northern Al-Mazar (Arabic: المزار الشمالي) is a town in the Irbid Governorate of north-western Jordan. Located 15 ...

 

Airport in Jamshedpur, Jharkhand Sonari AirportIATA: IXWICAO: VEJSSummaryAirport typePublicOwnerTata SteelOperatorAirports Authority of IndiaServesJamshedpurLocationSonari, Jamshedpur, Jharkhand, IndiaOpened1940; 83 years ago (1940)Elevation AMSL146 m / 478 ftCoordinates22°48′49″N 086°10′5″E / 22.81361°N 86.16806°E / 22.81361; 86.16806WebsiteSonari AirportMapIXWLocation of the airport in JharkhandShow map of JharkhandIXWIXW ...

Medieval castle in Seda, Alter do Chão, Portugal Castle of SedaCastelo de SedaPortalegre, Alto Alentejo, Alentejo in PortugalCoordinates39°11′35.163″N 7°47′13.221″W / 39.19310083°N 7.78700583°W / 39.19310083; -7.78700583TypeCastleSite informationOwnerPortuguese RepublicOpen tothe publicPublicSite historyMaterialsSchist, Limestone, Mortar The Castle of Seda (Portuguese: Castelo de Seda) is a medieval castle located in the civil parish of Seda...

 

Italian architect (1570–1655) Girolamo RainaldiA self portrait of Girolamo RainaldiBorn4 May 1570, RomeDied15 July 1655, RomeNationalityItalianKnown forRenaissance architecture Girolamo Rainaldi (4 May 1570 – 15 July 1655) was an Italian architect who worked mainly in a conservative Mannerist style, often with collaborating architects. He was a successful competitor of Bernini. His son, Carlo Rainaldi, became an even more notable, more fully Baroque architect. Biography The son of a ...

 

Samuel Xavier Datos personalesNombre completo Samuel Xavier BritoNacimiento São Paulo6 de junio de 1990 (33 años)País BrasilNacionalidad(es) BrasileñaAltura 1,67 m (5′ 6″)Carrera deportivaDeporte FútbolClub profesionalDebut deportivo 2010(Paulista)Club FluminenseLiga Serie APosición Lateral derechoDorsal(es) 2Trayectoria Paulista (2010-2012) São Caetano (2012-2013) Ceará (2014-2015) Sport Recife (2015-2018) → Atlético Mineiro (2018) → Ceará (2018) Ceará (2019-2021...

Penampang melintang lapisan ozon Bumi Lapisan ozon adalah lapisan di atmosfer pada ketinggian 20−35 km di atas permukaan Bumi yang mengandung molekul-molekul ozon.[1] Konsentrasi ozon di lapisan ini mencapai 10 ppm dan terbentuk akibat pengaruh sinar ultraviolet Matahari terhadap molekul-molekul oksigen. Peristiwa ini telah terjadi sejak berjuta-juta tahun yang lalu, tetapi campuran molekul-molekul nitrogen yang muncul di atmosfer menjaga konsentrasi ozon relatif stabil. Lapisa...

 

Szczepan Ścibior płk pil. Szczepan Ścibior pułkownik Data i miejsce urodzenia 13 grudnia 1903 Uniejów, Królestwo Polskie Data i miejsce śmierci 7 sierpnia 1952 Warszawa, Polska Przebieg służby Lata służby od 1924 Siły zbrojne Wojsko Polskie Polskie Siły Zbrojne Formacja Lotnictwo Wojska Polskiego Armée de l’air RAF Siły Powietrzne RP Jednostki 3 pułk lotniczydywizjon 305 Stanowiska komendant Oficerskiej Szkoły Lotniczej Wojska Polskiego w Dębliniedowódca eskadr...

 

Strategi Solo vs Squad di Free Fire: Cara Menang Mudah!